METHUSELAH ARCHIVE
SOURCES / FURRER-NEJM-ZAPPER-PACEMAKER-2004

Hazards of an alternative medicine device in a patient with a pacemaker

clinical paper · 2004
type:clinical paper
year:2004
citation:Furrer M, Naegeli B, Bertel O. Hazards of an alternative medicine device in a patient with a pacemaker. New England Journal of Medicine 350, no. 16 (April 15, 2004): 1688-1690. PMID: 15084709. DOI: 10.1056/NEJM200404153501623.
LINK
https://www.nejm.org/doi/full/10.1056/NEJM200404153501623
SUMMARY
Letter to the editor, Stadtspital Triemli, Zurich. Bibliographic fields copied from the NEJM article page fetched for this case, from Crossref for DOI 10.1056/NEJM200404153501623 (title, authors Marcel Furrer, Barbara Naegeli, and Osmund Bertel, journal, volume 350, issue 16, pages 1688-1690, issued April 15, 2004), and from PubMed esummary for PMID 15084709 (same title, authors, journal, volume, issue, pages 1688-90, publication date 2004 Apr 15). The letter reports a 52-year-old man with a dual-chamber pacemaker, implanted for intermittent high-degree atrioventricular block, who had short episodes of dizziness and near-fainting only during self-prescribed use of a Zapper. It describes the device as a battery-powered, direct-current-offset pulse generator with square-wave output at 33.3 kHz plus or minus 10 percent, applied to both hands, and as an alternative treatment used to eliminate cancer, other chronic illnesses, and self-diagnosed parasites and germs, citing Hulda R. Clark, The Cure for All Diseases (Chula Vista: New Century Press, 1995). Pacemaker settings were within limits; during use, high-frequency signals and ventricular oversensing with inhibition appeared, and the symptoms were reproduced, with a slow ventricular escape rhythm. The package carried a warning about pacemakers, which the patient disregarded. This is a single-patient device-interaction report, not a trial of cancer or parasite claims.
NOTES

One documented hazard: a zapper inhibited a pacemaker and the authors reproduced the patient’s near-fainting. It does not measure whether the device affects cancer or parasites.